Citation
Thiagarajah, Siva Priya and Jamil, Nur Akmar and Gan, Siew Mei and Khairulazdan, Nurkhairul Bariyah and Hamzah, Azrul Azlan and Majlis, Burhanuddin Yeop and Menon, P. Susthitha (2019) Detection of Uric Acid Using Kretschmann-based SPR Biosensor with MoS2-Graphene. In: 16th IEEE Student Conference on Research and Development, SCOReD 2018, 26 -28 November 2018, Selangor, Malaysia.
Text
152.pdf - Published Version Restricted to Repository staff only Download (448kB) |
Abstract
The paper investigates a Kretschmann-based surface plasmon resonance (SPR) biosensor based on graphene with high detection accuracy for uric acid sensing. Kretschmann arrangement is the most effective method for plasmon excitation. In this research, we examined the outcome of MoS 2 with a single layer of graphene deposited on top of gold thin film. Finite-difference-time-domain simulation technique is employed for investigation. The sensitivity and full-width-at-half-maximum (FWHM) of the SPR spectrum was analyzed to evaluate the performance of the SPR biosensor. The measurements are observed at visible and near-infrared wavelengths for uric acid detection. The sensitivity of the designed SPR sensor with a monolayer of MoS2 and graphene layer is achieved at 85.25°/RIU and 74.61°/RIU at 670 nm and 785 nm respectively. It is found that the proposed sensor has better detection accuracy at near-infrared wavelength with smaller FWHM compared to visible wavelength for detection of uric acid.
Item Type: | Conference or Workshop Item (Paper) |
---|---|
Uncontrolled Keywords: | Biosensors |
Subjects: | R Medicine > R Medicine (General) > R856-857 Biomedical engineering. Electronics. Instrumentation |
Divisions: | Faculty of Engineering (FOE) |
Depositing User: | Ms Suzilawati Abu Samah |
Date Deposited: | 26 Jan 2022 03:34 |
Last Modified: | 26 Jan 2022 03:34 |
URII: | http://shdl.mmu.edu.my/id/eprint/9025 |
Downloads
Downloads per month over past year
Edit (login required) |