Begum, Mahfuza and Almohammed, Huda I. and Rahman, A. K. M. Mizanur and Abdul Rashid, Hairul Azhar and Yusoff, Zulfadzli and Hamd, Zuhal Y. and Khandaker, Mayeen Uddin and Bradley, D. A. (2023) Collapsed photonic crystal fibre thermoluminescence dosimetry for external beam radiotherapy. Journal of Radiation Research and Applied Sciences, 16 (1). p. 100526. ISSN 1687-8507

[img] Text
41.pdf - Published Version
Restricted to Repository staff only

Download (2MB)

Abstract

In measuring exposure to ionizing radiation various detectors and techniques have been developed. Current investigation is of key TL dosimetric responses of Ge-doped collapsed photonic crystal fibre (PCF), a form offering the advantage of greater sensitivity compared to more conventional single-mode optical fibers. The PCF samples that have been studied herein have been used to obtain external photon beam percentage depth doses (PDD), elevated sensitivity being desirable in providing a basis for the capture of the range of doses. Using a linac operated at 6 and 10 MV, the results show a favourable comparison with ionization chamber values (typically to within 5%). Threshold dose, reproducibility, and dose-rate dependency have also been characterized.

Item Type: Article
Uncontrolled Keywords: Thermoluminescence, Percentage depth doses (PDD) Threshold dose Reproducibility Dose-rate dependency Ge-doped collapsed photonic crystal fibre (PCF) Solid water equivalent phantoms
Subjects: Q Science > QC Physics > QC474-496.9 Radiation physics (General)
Divisions: Faculty of Engineering (FOE)
Depositing User: Ms Nurul Iqtiani Ahmad
Date Deposited: 06 Jun 2023 02:15
Last Modified: 06 Jun 2023 02:15
URII: http://shdl.mmu.edu.my/id/eprint/11473

Downloads

Downloads per month over past year

View ItemEdit (login required)